Shutdown method in Empower does not stop the flow - WKB194091
Article number: 194091
SYMPTOMS
- The shutdown method in Empower does not stop the flow
ENVIRONMENT
- Breeze 2
- Empower 3
- Empower 2
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
CAUSE
The shutdown method event time to turn off flow is the same as the run time.
FIX or WORKAROUND
Adjust the run time or the event time to something different. They cannot be the same.