With Empower, Oracle's TEMP file grows very fast and continuously - WKB194003

Article number: 194003

SYMPTOMS

  • TEMP file grows to 32GB in one or two days
  • If another file is added, the second file also grows in one to two days

ENVIRONMENT

  • Empower 3 FR5 Enterprise
  • Vmware virtual machine
  • waters_connect

CAUSE

There was heavy swapping registered in the alert.log. RAM memory was set very low (8GB) and was shared (not dedicated) in the virtual machine.

FIX or WORKAROUND

  1. In the Vcenter, check the properties of the Virtual Machine and select "Reserve all guest memory".

    clipboard_e174b8c8c2c795143108caf290f3cd3bb.png
  2. Expand the RAM memory to 16 GB. In the standard installation, only Oracle is consuming more than 8 GB.