Agilent LC or AgGC status "No" in node properties with Agilent ICF Support 3.4 on Empower 3 - WKB193558
Article number: 193558
SYMPTOMS
- After upgrading to Agilent ICF Support 3.4, Agilent 1200 always appears with status "No" in node properties
- Can communicate, control, and run LC using Agilent TestApp
- Same issue with Agilent GC configured as AgGC (not issue if using legacy driver A7890)
- AgilentPlugInServer.exe process does not start in Task Manager, and when trying to start it manually, the following error occurs:
"AgilentPlugInServer.exe - System Error: The program can't start because api-ms-win-crt-runtime-l1-1-0.dll is missing from your computer. Try reinstalling the program to fix this problem."
ENVIRONMENT
- Empower 3 FR2 LAC/E
- Windows 7
- Agilent ICF Support 3.4
- Agilent ICF Support 3.5.1
CAUSE
KB2999226 is not installed on the LAC/E.
FIX or WORKAROUND
- Download KB2999226 from the Microsoft website (or reference this article) and install it on the LAC/E.
- Restart the LAC/E.
ADDITIONAL INFORMATION
KB2999226 was previously documented as a requirement in the Agilent ICF 3.0 release notes, but it is not documented in the Agilent ICF Support 3.4 release notes. This issue was recorded as CRI-3276.
Use Windows Local Administrator account to perform installation of Agilent ICF Support Layer