Intermittent communication problems with Agilent 7697 - WKB1934

Article number: 1934

SYMPTOMS

  • Communication problems with the HSS 7697

ENVIRONMENT

  • Empower
  • Headspace
  • HCO 3.0
  • Firmware A.01.06.1

CAUSE

Enabling the DHCP on the 7697 is not supported.

FIX or WORKAROUND

  1. Set a static IP address on the 7697.
  2. Run as normal.

ADDITIONAL INFORMATION

Enhancement Request was submitted to test DHCP enabling on the next HCO revision PCS#55151.

Headspace Control Option Version 3.0

See article WKB97285 on Headspace Sample preparation