Peaks are not being identified in Empower 3 - WKB18855

Article number: 18855

SYMPTOMS

  • After integrating and quantitating the data, two out of three peaks are not identified in the Review window in Empower 3

ENVIRONMENT

  • Empower 3
  • Old processing method that worked in the past
  • Peak response is lower than in previous runs

CAUSE

The peak threshold is set too high, causing the peaks with a lower response to not be integrated.

FIX or WORKAROUND

Set the peak threshold low enough so that the other two peaks are integrated.

ADDITIONAL INFORMATION

 