"Injection error - Missing/wrong plate or vial" when using 96 Well Plates 186005837 (700ul) and 186002481 (800ul) - WKB1881
Article number: 1881
SYMPTOMS
injection error - Missing/wrong plate or vial
ENVIRONMENT
- Empower
- MassLynx
- ACQUITY I-Class SM-FTN
- ACQUITY H-Class SM-FTN
- ACQUITY UPLC SM
- ACQUITY I-Class SM-FL
- ACQUITY M-Class SM-FL
- nanoACQUITY SM
- ACQUITY UPC2SM-FL
- 186005837 Plate, 96-Well, 700 uL, Round Well, 25/PK
- 186002481 1-mL Round Collection Plate, 50/PK
CAUSE
Selecting the wrong plate type in the Empower sample set or MassLynx bed layout triggers this error
FIX or WORKAROUND
- Select "ANSI-96-well 1 mL" as the plate type for part numbers 186005837 and 186002481 in Empower Sample Set or MassLynx Bed Layout
- Reference Waters Chromatography Columns and Supplies Catalog for more information
ADDITIONAL INFORMATION
- Confirm plate dimensions in MassLynx: Select Autosampler > Configure > Plate Generator
- Confirm plate dimensions in Empower: Configure System > Plate Types > Select Plate > right-click Properties

