Error seen in Empower message center: Instrument Failure e-SAT/IN - WKB17948
Article number: 17948
SYMPTOMS
- The socket failed while waiting for incoming data; error code WinSock Error # 0
- A failure occurred in reading the reply of Clear instrument
ENVIRONMENT
- Empower 3 FR2
- e-SAT/IN
CAUSE
Two IP address assignments in the same subnet for the Instrument LAN
FIX or WORKAROUND
Remove one of the IP address assignments.
ADDITIONAL INFORMATION