Unable to restore an Empower 2 project into Empower 3 - WKB17927
Article number: 17927
SYMPTOMS
- Error message when attempting to restore: "Project version - 7.21.00.00 Current version - 6.21.00.00, Would you like to attempt to update this project?"
- Allowing attempt to update yields "Update of project failed"
- When restoring a project from Empower 2 into Empower 3, the update process fails and deletes the project from Empower 3
ENVIRONMENT
- Empower 2
- Including FR5 HF1
- Empower 3
- Including FR2 SR2 HF1
CAUSE
There are many possible causes of this error, but a common cause is a blank value for the injection vial number.
FIX or WORKAROUND
- Restore the project back into Empower 2.
- Open the project and select the Injection tab.
- If there are more than 1000 injections, modify the display from 1000 to 10000.
- Click on the vial column to sort the list.
- Vial 1 should be the top vial. If there are any blank injections, delete them.
- Back up the project and restore it into Empower 3.