Duplicate results for a Result Set processed once - WKB1553
Article number: 1553
SYMPTOMS
- Standards look fine.
- Two sets of data for unknown.
- Samples are in duplicate, data looks identical.
- More Results than expected for a Result Set processed once.
ENVIRONMENT
- Empower 2
- Empower 3
CAUSE
The sample set is processing the unknowns twice. Once, because the processing column is set to "normal", and then they are processed again when it hits the Quantitate line in the sample set.

FIX or WORKAROUND
- For the unknowns, set the processing column to "Don't Process or Report". It will then only process when it gets to the Quantitate line.