Injections are not included when using Data Filtering Conditions for a report group in Empower - WKB15480

Article number: 15480

SYMPTOMS

  • Injections are not included when using Data Filtering Conditions for a report group in Empower
  • Data filtering conditions included sample type, component type, and label; sample type filtered to be just "Unknown", component type filtered to be "Main Component", and labels to include "u*" and "U*"
  • With the data filtering conditions below, it did not filter out components that are not main components

  clipboard_e15ade12978d3e8ae293a2a9fe655771b.png

clipboard_ed90f2d277093c997cff7b43f797819d3.png

ENVIRONMENT

  • Empower 3 Feature Release 3

CAUSE

Data filtering conditions were not configured properly.

FIX or WORKAROUND

  1. For data filtering conditions, each line is an "OR" and each column is an "AND". If you want multiple conditions, such as including label, component name, and sample type for multiple scenarios, each condition corresponds to one condition. Below is an example:

clipboard_ef58c2673350382aeca5df4ea67798122.png

clipboard_eb35b6e71a3d4ea1072fce67f94b79b07.png

 

ADDITIONAL INFORMATION

 

 