Error "Project name failed to copy, could not be updated to the current database version" observed in Empower - WKB14604
Article number: 14604
SYMPTOMS
- Error observed when restoring project
- Works when a domain administrator is logged on to the Windows account
- Does not work when a regular Windows domain user is logged in
ENVIRONMENT
- Empower 2 FR5 Enterprise
CAUSE
Current Windows account does not have the correct permissions for the raw data or target folders.
FIX or WORKAROUND
- Add authenticated users to the sharing and security permissions tabs of the raw data share folder properties .