Standards show no amounts in Breeze - WKB1412

Article number: 1412

SYMPTOMS

When standard injections are processed with the same processing method as always the results show no amounts

ENVIRONMENT

  • Breeze

CAUSE

Injected Standards as Sample Type - Unknown

No amounts entered

FIX or WORKAROUND

  1. Click on Find data after logging to Breeze ( username is Breeze, no password)
  2. Click on the Injections tab
  3. Highlight injection , right click and click on Alter Sample
  4. Change Sample Type from Unknown to Standards
  5. File then click Save
  6. Click on Edit then click on Amounts
  7. Enter Amounts, then click File - Save ( alternatively you can edit Processing Method and enter amounts under "default amounts")
  8. Reprocess and save results

ADDITIONAL INFORMATION