Skip to main content
Waters

Carrier gas flow drops with an Agilent 7697 Headspace using Empower - WKB14107

Article number: 14107

SYMPTOMS

  • Carrier gas flow drops as soon as the 6890 GC and 7697 Headspace are registered under Empower control
  • Choosing Headspace control, GC control, or GC and Headspace control for flow does not resolve the issue
  • The 7697 Headspace has the optional carrier EPC. A second 7697 Headspace does not have the optional carrier EPC and does not drop carrier gas flow when controlled by Empower

ENVIRONMENT

  • Agilent 7697 Headspace
  • Agilent 6890 GC
  • Agilent 7890 GC
  • Headspace Control Option (HCO) 3.0
  • Empower 3

CAUSE

The optional carrier EPC ("flow control chip") for the 7697 is not supported. This is a known issue. GC control of carrier gas flow is supported.

FIX or WORKAROUND

The optional carrier EPC is supported for the 7697 in Instrument Control Framework (ICF) v3.0.

ADDITIONAL INFORMATION

This also occurs when using 7890 GC with 7697A Headspace.

When configuring the GC/HS system with ICF 3.0, there is an option "instrument controlling carrier setpoints". Use this section to specify which instrument controls the carrier gas flow. If the HS includes an optional EPC module for carrier gas flow, all options will be available. If the HS does not have the optional EPC, the setting is GC instrument.

clipboard_ef98533218c11f857dd44f6ce965b3593.png

See article WKB97285 on Headspace Sample preparation

Not able to find a solution? Click here to request help.