2414 RI and e2695: Sample Set pauses at Purge Detector line - WKB13498

Article number: 13498

SYMPTOMS

  • Sample Set pauses at the Purge Detector line and does not proceed to the next line in the Sample Set

ENVIRONMENT

  • Empower 3
  • 2414 RI detector
  • e2695 Alliance

CAUSE

PCS 51061 - For systems that include a 2414 detector, when you initiate a sample set that contains a Purge Detector function, and then close the Run Samples window, the sample set pauses after the Purge Detector function completes (see  PCS  51061 )

FIX or WORKAROUND

The Sample Set run resumes when you reopen Run Samples. Alternatively, to avoid this issue, do not close the Run Samples window when you have a Purge Detector line in the Sample Set.

If possible , configure your system to include only the e2695 and 2414 detector

ADDITIONAL INFORMATION

This is related to the firmware version.

  • With 2.04 once under Empower control the 2414 switches the mode to 410.
  • With 2.10 with the 2414 ICS loaded this does not happen, it stays in RIU.

 

Same issue if instead of Purge Detector there is a Purge Injector function.

 