Agilent 6890 GC does not communicate in Empower - WKB13490
Article number: 13490
SYMPTOMS
- Agilent 6890 GC does not connect Instrument Status OK = "No" Serial port is configured correctly ommunication settings on the GC are configured correctly.
ENVIRONMENT
- Empower 3 FR2 LAC/E Direct connection to the COM port Agilent 6890 GC.
CAUSE
Bad cable.
FIX or WORKAROUND
- Replace cable: 441001561, Ser Cbl 15 ft 9-pin F/F Null Modem RoHS.
ADDITIONAL INFORMATION
To order this part, call your local Customer Service department.