Citrix server hangs; cannot open remote desktop to launch Empower - WKB12950

Article number: 12950

SYMPTOMS

  • Login to Citrix server via Remote Desktop produces the "Welcome" screen, where it remains indefinitely
  • Launching the Empower published application from a thin client does not launch Empower
  • Existing connections/users on the Citrix server are running Empower without issues
  • New connections cannot be established

ENVIRONMENT

  • Citrix Xenapp 6.5
  • Windows Server 2008 R2
  • Empower 3 SR2

CAUSE

  • Microsoft Support's analysis of the dump file identified the root cause of the issue as an empower.exe process hangup
  • Further analysis identified that emdriver.sys was tied to that empower.exe process
  • emdriver.sys is part of the third-party software utility AppSense
  • AppSense causes the empower.exe process to hang

FIX or WORKAROUND

Remove/disable any interactions/links between the Empower software and the AppSense application.

(The customer's AppSense administrators created scripts to remove Empower applications from the AppSense scope.)

ADDITIONAL INFORMATION

 