In Empower 3, how do you set acceptance criteria for test results without using custom fields? - WKB122994
Article number: 122994
OBJECTIVE or GOAL
In Empower 3, set acceptance criteria for test results with reporting methods, without using custom fields.
ENVIRONMENT
- Empower 3
- Empower 2
PROCEDURE
- Open Table Properties in the Report Method Editor window.
- Set the required columns for the table, and on the Table tab, select the column for which you want to set the acceptance criteria. (Click the column name to highlight it with black.)
- Select the Column Properties button to open the Column Properties dialog box.
- Set the function you want to evaluate in the summary function table, and enter a criterion value for the minimum or maximum value.

5. In the table shown for reference, the maximum value of %RSD of the area value is set to 0.2000, resulting in values exceeding the value displayed in red italics.