Empower 3 - Sample set stuck in "setting up" - WKB114449
Article number: 114449
SYMPTOMS
- Sample set stuck in "setting up"
- Sample set runs the first injection of the set, but does not move on to the second injection
- The injection timer continues counting up past the method end time and does not progress to the second injection
ENVIRONMENT
- Empower 3
CAUSE
An additional detector is in the instrument configuration but is not part of the instrument method.
FIX or WORKAROUND
- Remove from the instrument configuration any modules that are not being used.
- Configuration manager > System > right-click "system properties".
- Drag any unused components to the left-hand side and click OK.