Sample set does not start after pressing the Run button - WKB112916
Article number: 112916
SYMPTOMS
- System will go to waiting for injection
- Autosampler does not start injection cycle
ENVIRONMENT
- Empower 3
- Waters Fraction Manager
CAUSE
Instrument configuration does not match instrument method.
FIX or WORKAROUND
Recreate the system to match the instrument method.
ADDITIONAL INFORMATION
WFM - does not appear to be considered when allowing instrument methods to be displayed for a system.