TEMP data file created in user's TEMP directory may grow fast enough to fill drive space in LAC/E - WKB112689
Article number: 112689
SYMPTOMS
- Error in the message center: “Instrument failure Bad data”
- Error in the message center: “Not enough memory resources are available to process this command” followed by “Chrom map view of file failed”
- Server Busy in the Run Samples window
ENVIRONMENT
- Empower 3 Enterprise
- Agilent Instrument Control Framework v3.2
- Agilent LC
- Agilent GC
CAUSE
Previous failure status of the Agilentpluginserver.exe can lead to this behavior.
FIX or WORKAROUND
-
If a redundant file with the name “tempdatax” (where x is the file number of that channel) is seen on the user's TEMP directory, delete the file.
By default, the location of this file is C:\Users\AppData\Local\Temp directory. Otherwise, this temp directory may be at a location configured by their IT administrator for the specific PC.
2. This issue is fixed in ICF 3.3 as part of fixing the "Instrument failure Bad data" error.
ADDITIONAL INFORMATION
CRI - 2548
Service Alert - 715007115
id112689, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP