Raw data (.dat) file fails to copy in Empower during project restore - WKB11266
Article number: 11266
SYMPTOMS
- Error message " D4710.dat is failing to copy " during project restore from a backup
ENVIRONMENT
- Empower 3 Feature Release 3
CAUSE
Dxxxx.dat file size is less than 1 kB.
FIX or WORKAROUND
- Delete the file and proceed with restore.