Caffeine amount not calculating in Breeze running the Soft Drink Analysis - WKB11258
Article number: 11258
SYMPTOMS
- All peaks are present
- Retention time is within the RT window
- Caffeine amount is not calculating
- Aspartame, benzoate, and sorbate all have amounts
ENVIRONMENT
- Breeze 2
- New column
- Soft Drink Analysis
CAUSE
Inhibit integration ended after the peak started to elute.
FIX or WORKAROUND
- Change the inhibit integration to an earlier time.