Ethernet communication issues; Node Properties "No" - WKB108266
- Last updated
-
-
Save as PDF
-
<h2>SYMPTOMS</h2>
<ul>
<li class="body-text">Some system components show "no" in node properties</li>
<li class="body-text">All instruments appear in DHCP </li>
<li class="body-text">All instruments can be pinged</li>
</ul>
<h2>ENVIRONMENT</h2>
<ul>
<li class="body-text">Empower 3</li>
</ul>
<h2>CAUSE</h2>
<p>Corrupted ICS</p>
<h2>FIX or WORKAROUND</h2>
<ol>
<li class="body-text">Unload the current ICS for the affected component.</li>
<li class="body-text">Reload ICS.</li>
<li class="body-text">Rerun the DHCP Server Configuration Wizard.</li>
</ol>