Skip to main content
Waters

Retention time differs between a single wavelength and an extracted single wavelength from the same injection - WKB10308

Article number: 10308

SYMPTOMS

  • A 3D channel with an extracted wavelength and a 2D channel of the same wavelength have different retention times for their peaks

ENVIRONMENT

  • Xcalibur 
  • ACQUITY UPLC PDA eLambda detector 

CAUSE

The 3D scan and the 2D channel use different resolution values in the instrument method.

FIX or WORKAROUND

  1. Change the resolution values so they match.
  2.  Rerun.

ADDITIONAL INFORMATION

Example chromatogram

clipboard_ea494e467acd6acbff1c5317d919be510.png

Not able to find a solution? Click here to request help.