SYMPTOMS

  • Sample set processes, but no result set or results are created
  • Results can be created if processed in Review
  • Review error messages: "Result was not saved" and "Unable to save all results"

ENVIRONMENT

  • Empower 3 FR4

CAUSE

Low tablespace (<9 Mb)

FIX or WORKAROUND

Increase tablespace by 25 Mb.