Result set is not created after processing a sample set - WKB101556
Article number: 101556
SYMPTOMS
- Sample set processes, but no result set or results are created
- Results can be created if processed in Review
- Review error messages: "Result was not saved" and "Unable to save all results"
ENVIRONMENT
- Empower 3 FR4
CAUSE
Low tablespace (<9 Mb)
FIX or WORKAROUND
Increase tablespace by 25 Mb.
ADDITIONAL INFORMATION