Data is buffering on one LAC/E32 but is not sent to the server - WKB101538
Article number: 101538
SYMPTOMS
- After acquisition, data is still sitting on the LAC/E box and can be seen in the project folders
- Per recover.log file, data is buffering on the LAC/E
- Only one LAC/E in the network is having this issue
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
WatersService is disabled.
FIX or WORKAROUND
Enable the WatersService and start it. Data will recover in 10 minutes.