Alliance e2695 does not communicate with Empower 3 - WKB10011
Article number: 10011
SYMPTOMS
- e2695 Alliance Separations Module is not seen in the node properties
- Does not populate in the DHCP Configuration
- Does not respond to a ping
- Ethernet LEDs on the instrument are not lit
ENVIRONMENT
- Empower 3 FR3
- e2695 Alliance Separations Module
CAUSE
The e2695 was not configured for Ethernet communication.
FIX or WORKAROUND
- From the main menu on the front panel of the Alliance module, select Config.
- Navigate to the Network Address field and select Auto/Ethernet.
- Restart the Alliance module.