After removing it, the Agilent G1888 is still present in the node properties of Empower - WKB10005
Article number: 10005
SYMPTOMS
- After removing it, the Agilent G1888 still shows in the Empower node properties
ENVIRONMENT
- Empower 3
- Headspace Control Option v3.0
- Agilent GC with G1888 Headspace
- Static IP address
CAUSE
The G1888 was manually added to the DHCP, so it will always be present unless you remove it.
FIX or WORKAROUND
- Remove the Agilent G1888 from the DHCP configuration.
- Remove the Agilent G1888 from the node properties.
- Reboot LACE PC.