What are possible reasons for low recovery of histidine after acid hydrolysis of a pure protein API? - WKB190630

Article number: 190630

ENVIRONMENT

  • Amino acid analysis (AAA)
  • Histidine
  • Acid hydrolysis
  • AccQ•Tag
  • Pico•Tag

ANSWER

  • Constant boiling phenol was not added to the hydrolysis tubes.
  • Histidine cleavage is much slower than other amino acids and can be inhibited by amine-containing buffers like TRIS. Prior to hydrolysis, ensure that the protein is free from amine-containing buffers. 
  • An internal standard is not being used.

ADDITIONAL INFORMATION

For additional information:

Does waters have a recommended protein hydrolysis method for AccQ.Tag? 

When should the internal standard be added in amino acid analysis?

Fountoulakis, Michael, Lahm, Hans-Werner, "Hydrolysis and amino acid composition analysis of proteins," Journal of Chromatography A, 826, 109-134 (1998).