Can I clean my ACQUITY APC column in reverse direction? - WKB81352
Article number: 81352
ENVIRONMENT
- Advanced polymer chromatography (APC)
- ACQUITY APC columns
- Column cleaning
ANSWER
Yes, but only if cleaning in the standard direction does not bring the column pressure down to remove a clog or precipitation.
ADDITIONAL INFORMATION
The column flow rate must be slowed to 0.1 mL/min to clean in the reverse direction.
For additional cleaning suggestions, see Section V, Cleaning and Regeneration in the ACQUITY APC Columns Care and Use Manual.
id81352,
