Can I clean my ACQUITY APC column in reverse direction? - WKB81352

Article number: 81352

ENVIRONMENT

  • Advanced polymer chromatography (APC)
  • ACQUITY APC columns
  • Column cleaning

ANSWER

Yes, but only if cleaning in the standard direction does not bring the column pressure down to remove a clog or precipitation.

ADDITIONAL INFORMATION

The column flow rate must be slowed to 0.1 mL/min to clean in the reverse direction.  

For additional cleaning suggestions, see Section V, Cleaning and Regeneration in the ACQUITY APC Columns Care and Use Manual.