How do I increase retention time for a polar compound on a liquid chromatography (LC) column? - WKB74280
Article number: 74280
OBJECTIVE or GOAL
Increase retention time of polar compounds in reverse-phase mode.
ENVIRONMENT
- Polar compunds
- Reverse phase
- Method optimization
- Method development
- HSS T3
- CORTECS T3
PROCEDURE
- Equilibrate the column for at least 20 column volumes.
- Lower the percentage of organic in the mobile phase, or run 100% aqueous mobile phase to help promote retention of polar analytes.
- Slow the flow rate of the method to promote more contact time with the sorbent.
- If the polar compound is ionizable, add a mobile phase modifier that will change the pH to promote retention.
- Change column chemistry to a different reverse-phase chemistry.
- Explore alternative retention modes for polar analytes.
ADDITIONAL INFORMATION
For additional information, see Which C18 column will best retain polar analytes with reverse phase? and What is a good column choice for the retention of polar, small molecules that are zwitterionic? and What is one of the causes of poor retention of an analyte on a column?