What is a good cleaning procedure for the soft drink analysis column? - WKB74185

Article number: 74185

ENVIRONMENT

  • Column cleaning
  • Soft drink analysis
  • Nova-Pak C18

ANSWER

The column can be cleaned with 80:20 methanol/water for 50 minutes at 1 mL/min. It can be flushed with 100% acetonitrile for additional 35 minutes at 1 mL/min.  

ADDITIONAL INFORMATION