Skip to main content
Waters

Can I run 100 percent aqueous mobile phase through XBridge C18 columns? - WKB27769

Article number: 27769

ENVIRONMENT

  • XBridge columns
  • Dewetting

ANSWER

Yes, you can run 100% aqueous phase through an XBridge C18. If you shut the system down, it is recommended that organic be introduced through the column to help rewet the pore prior to running 100% aqueous.  

ADDITIONAL INFORMATION

The XBridge BEH C18 must not sit in or be stored in 100% aqueous mobile phases. It should be flushed with at least 20% organic every day. If it will not be used for four days or longer, it must be stored in 100% acetonitrile.

For additional information, see the XBridge Columns Care and Use Manual.

id27769, eluent