What are best practices to consider in analysis of oligonucleotides by ion-pair reverse-phase separations? - WKB276724
Article number: 276724
ENVIRONMENT
- Oligonucleotide analysis
- OST (oligonucleotide separation technology)
- Column characteristics
- best practices
- Column care and use
- Reverse phase
- C18
- Chromatography
ADDITIONAL INFORMATION
Avoid the use of glass mobile phase container filters/sinkers for the high-pH mobile phases. The reverse-phase oligo application typically runs at pH >7, which is ideal for the oligos but not for glass mobile phase container filters/sinkers; the high pH dissolves the borosilicate filter fibers faster. Dissolved silicate particles can clog columns and change the surface chemistry of the columns.
Use titanium mobile phase container filters or high-quality Waters Critical Clean Stainless Steel mobile phase filters.
See also What is the column cleaning procedure to reduce sodium adducts in oligonucleotide LC/MS analysis?
