What is the difference between XBridge Premier Protein SEC 250A, XBridge BEH SEC 200A, and BioResolve SEC mAb columns? - WKB252444
Article number: 252444
ENVIRONMENT
- Column comparison
- ACQUITY
- XBridge
- Bridged ethylene hybrid (BEH)
- Size exclusion chromatography (SEC)
- High Performance Surfaces (HPS)
- Polyethylene oxide (PEO)
- MaxPeak Premier
ANSWER
XBridge Premier Protein SEC 250A columns:
- 250A pore size
- PEO-bonded BEH particles (prevent protein adsorption to the particle)
- 2.5-µm particle size
- MaxPeak High Performance Surfaces hardware (prevents protein adsorption to the stainless steel frits or column body)
- First choice for best performance
XBridge BEH SEC 200A columns:
- 200A pore size
- Diol-bonded BEH particles
- 2.5-µm and 3.5-µm particle sizes
- Traditional stainless steel hardware
- Require conditioning injections in order to achieve best recovery of proteins
BioResolve SEC mAb columns:
- 200A pore size
- Diol-bonded BEH particles
- 2.5-µm particle size only
- Traditional stainless steel hardware
- Should not require conditioning injections in most cases
ADDITIONAL INFORMATION
MaxPeak Premier Protein SEC Columns Infographic
BEH SEC Column Technology – Infographic
BioResolve SEC mAb Consumables- Infographic