How to clean a µBondapak C18 column - WKB252273

Article number: 252273

OBJECTIVE or GOAL

Clean a µBondapak C18 column.

ENVIRONMENT

  • Column care and use
  • Column cleaning
  • µBondapak C18

PROCEDURE

  1. Remove the guard column if one is in use.
  2. Remove the pre-column filter if one is in use.
  3. Disconnect the column outlet from your detector.
  4. Flow the outlet of the column to waste (bypass detector).
  5. Prepare a cleaning solution of 50:50 dimethylsulfoxide/water.
  6. Run a mobile phase of 100% water.
  7. Make repeated injections of the 50:50 dimethylsulfoxide/water cleaning solution every five minutes.
  8. Repeat step 7 until ten empty column volumes of mobile phase have passed through the column.
  9. After the cleaning injections are made:
    • Flush the column with 10 empty column volumes of water.
    • Flush the column with 10 empty column volumes of methanol.
    • Flush the column with 10 empty column volumes of chloroform.
    • Flush the column with 10 empty column volumes of methanol.
    • Flush the column with 10 empty column volumes of water.