What are the recommended gradient conditions in cleaning the ACQUITY BEH Glycan columns? - WKB121901
Article number: 121901
ENVIRONMENT
- ACQUITY BEH Glycan
- Column cleaning conditions
- Column care and use
ANSWER
The cleaning gradient may be 0-100% water over five column volumes (minimum) at 0.25 mL/min. Do three to five repetitions of the gradient to clean.
ADDITIONAL INFORMATION
For additional information, see How do I calculate the empty column volume of a column? and the ACQUITY UPLC BEH Glycan, 1.7 µm Columns and Glycan Performance Test Standard.
