How do I calculate the empty column volume of a column? - WKB48971
Article number: 48971
OBJECTIVE or GOAL
Determine the volume of an empty column.
ENVIRONMENT
- Column
- calculate
- empty volume
- HPLC
- UPLC
- ACQUITY
- HSS
- Symmetry
- BEH
- OST
- APC
- XBridge
PROCEDURE
- Use Volume = pi x radius2 x Length
- pi = 3.14
- r2 and Length should be converted to centimeters
- Diameter of column divided by 2 = radius
- radius x radius = r2
- 3.14 x r2 x L = Volume in cm3
- cm3 = 1 mL
ADDITIONAL INFORMATION
The empty column volume of 4.6 mm x 150 mm is 2.5 mL.
pi = 3.14
4.6 mm becomes 0.46 cm and 150 mm becomes 15 cm.
0.46 cm divided by 2 = 0.23 cm radius
0.23 cm x 0.23 cm = 0.0529 cm2
3.14 x 0.0529 cm2 x 15.0 cm = 2.49cm3
Also see --> How do I determine column void volume?