How should I clean an ACQUITY BEH SEC column or XBridge BEH SEC column if I suspect protein has precipitated on the column? - WKB108019
Article number: 108019
ENVIRONMENT
- Ultra Performance Liquid Chromatography (UPLC)
- Ultra High Performance Liquid Chromatography (UHPLC)
- High Performance Liquid Chromatography (HPLC)
- Bridged ethylene hybrid (BEH)
- Size exclusion chromatography (SEC)
- Column
- XBridge BEH SEC column
- ACQUITY BEH SEC column
- Protein separation
- Column cleaning
ANSWER
- Perform a shallow gradient from 0% to 100% acetonitrile over a period of 30 minutes, and perform this two times.
- Inject a protein molecular weight standard to verify that step 1 corrects the issue.
- If the issue is not corrected, choose one of the following options:
- Make three 100-µL injections 50:50 of DMSO:water (dimethylsulfoxide) if your system is compatible with DMSO.
- Make three 100-µL injections of 6M guanidnium chloride in a mobile phase of 50:50 H2O/acetonitrile.
ADDITIONAL INFORMATION
Prior to cleaning the column, inject a protein molecular weight standard to confirm that the column is fouled.
Also see:
How do I clean an ACQUITY UPLC BEH SEC column or an XBridge BEH SEC column?
Can I clean an XBridge BEH SEC Column or ACQUITY UPLC BEH SEC column with Guanidine or Urea?