Can I clean XBridge BEH SEC columns and ACQUITY UPLC BEH SEC Columns with injections of DMSO (dimethylsulfoxide)? - WKB50971
Article number: 50971
ENVIRONMENT
- XBridge BEH SEC Columns
- ACQUITY UPLC BEH SEC Columns
- Column Cleaning
ANSWER
Yes. See details in "additional information" section below.
ADDITIONAL INFORMATION
If cleaning the column with DMSO, the recommended cleaning injection solution should be 50:50 DMSO:H2O in a sample vial.
Then, run 100% H2O as the mobile phase and make 5 to 10 injections of the 50:50 DMSO:H2O solution as the cleaning solution.
The Care & Use Manuals provide other cleaning suggestions:
ACQUITY UPLC® BEH SEC Columns and Standards Care & Use Manual (lit code 720003385en): https://help.waters.com/help/en/support/library-details.html?documentid=720003385en
XBridge Protein BEH SEC Columns & Standards Care & Use Manual (lit code 720005206en): https://help.waters.com/help/en/support/library-details.html?documentid=720005206en