2996 PDA detector fails linearity test - WKB236600

Article number: 236600

SYMPTOMS

  • 2996 PDA Detector fails linearity test

ENVIRONMENT

  • Alliance 2695
  • 2996 PDA Detector

CAUSE

Diluted caffeine solutions from stock solution were used. Multiple failures occurred due to one or more calibration points being above or below the calibration line. After linearity test with certified Waters UV solutions, test is successful.

FIX or WORKAROUND

Use certified Waters caffeine solutions to ensure that solution concentrations are correct.

ADDITIONAL INFORMATION

HPLC System UV Test Solutions part number: 700003741

id236600, 2996, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP