Opening the source isolation valve
Article number: 79752
To open the source isolation valve after completing a maintenance procedure:
Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid puncture wounds, take great care while working with the source enclosure open if one or both of these conditions apply:
- An ESI probe is fitted (the probe’s tip is sharp).
- A corona pin is fitted (the pin’s tip is sharp).
- Open the source isolation valve by moving its handle clockwise to the horizontal position.
- Close the source enclosure.