Opening the source isolation valve

Article number: 79752

To open the source isolation valve after completing a maintenance procedure:
bio hazzard.PNG

Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.
Puncture wound.PNG 

Warning: To avoid puncture wounds, take great care while working with the source enclosure open if one or both of these conditions apply:

  • An ESI probe is fitted (the probe’s tip is sharp).
  • A corona pin is fitted (the pin’s tip is sharp).
  1. Open the source isolation valve by moving its handle clockwise to the horizontal position.

clipboard_ec09c40a6bc7f76fa141f1aa2e53bea7a.png

  1. Close the source enclosure.