Fitting the source enclosure to the instrument
Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument:
Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid puncture wounds, take great care while fitting the source enclosure to the source if a corona pin is fitted. The pin's tip is sharp.
- Using both hands, fit the source enclosure to the two supporting studs on the source adaptor housing.
Notice: To avoid damaging the sample inlet, when removing a NanoLockSpray source enclosure, you must slide the sprayer platform out of the source enclosure before you open the enclosure.
- Close the source enclosure.
- Connect the probe adjuster and options cables to the instrument’s connectors.
Tip: The cables and connectors are color coded; the blue-sleeved cable connects to the blue connector and the yellow-sleeved cable to the yellow connector.
- Slide closed the instrument’s source interface door.