Fitting the ESI probe tip and gasket

Article number: 76185

Required materials

  • Chemical-resistant, powder-free gloves
  • 10-mm open-end wrench
  • 7-mm open-end wrench
  • New metal gasket

To fit the ESI probe tip and gasket:
Hazzard.PNG Warning:
To avoid personal contamination with biologically hazardous materials, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Puncture wound.PNG Warning: To avoid puncture injuries, handle sample needles, syringes, fused silica lines, and borosilicate tips with extreme care.
exclamation mark.PNGNotice: To avoid damage, do not reuse metal gaskets. Always fit a new gasket.

  1. Fit the new metal gasket into the probe tip.


Inserting the gasket.PNG

  1. Fit the probe tip and screw the tip onto the probe assembly.
  2. Use the 7-mm wrench in conjunction with the 10-mm wrench to tighten the probe tip.

Important: To avoid gas leakage, fully tighten the probe tip.

  1. Fit the ESI probe adapter to the source (see Installing the probe adapter).
  2. If required, re-optimize the probe position.