Fitting the ESI probe tip and gasket
Article number: 76185
Required materials
- Chemical-resistant, powder-free gloves
- 10-mm open-end wrench
- 7-mm open-end wrench
- New metal gasket
To fit the ESI probe tip and gasket:
Warning: To avoid personal contamination with biologically hazardous materials, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid puncture injuries, handle sample needles, syringes, fused silica lines, and borosilicate tips with extreme care.
Notice: To avoid damage, do not reuse metal gaskets. Always fit a new gasket.
- Fit the new metal gasket into the probe tip.
- Fit the probe tip and screw the tip onto the probe assembly.
- Use the 7-mm wrench in conjunction with the 10-mm wrench to tighten the probe tip.
Important: To avoid gas leakage, fully tighten the probe tip.
- Fit the ESI probe adapter to the source (see Installing the probe adapter).
- If required, re-optimize the probe position.