Closing the source isolation valve

Article number: 79726

To close the source isolation valve before starting a maintenance procedure:
bio hazzard.PNG

Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.

  1. Prepare the instrument for working on the source (see Preparing the instrument for working on the source).

Burn.PNG Warning: To avoid burn injuries, take great care while working with the source enclosure open.
Puncture wound.PNG Warning: To avoid puncture wounds, take great care while working with the source enclosure open if one or both of these conditions apply:

  • An ESI probe is fitted (the probe’s tip is sharp).
  • A corona pin is fitted (the pin’s tip is sharp).
  1. Pull the source enclosure release (located at the bottom, right-hand side) outward and swing open the enclosure.
  2. Close the source isolation valve by turning its handle counterclockwise to the vertical position.