Before you begin using the instrument

Article number: 93380

Waters designed the SELECT SERIES cIMS instrument to be used as a research tool to deliver authenticated, exact-mass measurement. It is not for use in diagnostic procedures.

Requirements: The SELECT SERIES cIMS has been configured for your application by the Waters installer.

To prepare for data acquisition:

Before starting the instrument, you must complete the following tasks:

  • Ensure that the instrument is pumped down (evacuated) and the ionization source for your application is fitted.
  • Visually inspect the ion probe position.
  • Ensure that the wash bottle contains a sufficient volume of a suitable wash solution.

Tip: The composition of the wash solution is sample-dependent. However, for most applications, Waters recommends an 80:20 methanol/water mixture.

  • Ensure that your sample, lock-mass reference, and calibrant solutions are ready and mounted on the instrument.

Recommendation: Use reservoir bottles in this order: A for sample, B for Lockmass reference, C for calibrant.

Tip: Filter or centrifuge the sample solution to remove particulates before placing it in reservoir A, to avoid blocking the transfer line.

  • Ensure the waste line is correctly plumbed.
  • Ensure the LC system is configured and ready to use.

Not able to find a solution? Click here to request help.