How to install, setup and maintain the SELECT SERIES MALDI setup source
Article number: 243611
To insert and clean the SELECT SERIES MALDI setup source:
- Remove and clean the hexapole (see How to remove and clean the SELECT SERIES MALDI hexapole for more information).
- Insert the SELECT SERIES MALDI setup source using the guide rail.
- With the source pressed tightly against the O-ring, tighten the screws until they are finger tight.
- Connect the gas line, source ID and RF connection.
- Connect the sample line to the syringe drive.
- Select Pump on the tune page.
- Open the isolation valve.
Note: The source is now functioning.
To remove the source and re-insert the hexapole:
- Close the isolation valve
- Vent the SELECT SERIES MALDI setup source
- To remove the SELECT SERIES MALDI setup source, disconnect the gas line, source ID, RF connection, and the sample line.
- Clean the SELECT SERIES MALDI setup with nitrogen and store it.
- Re-insert the hexapole using the guide rail.