REM Method Editor
Article number: 269359
To acquire sample data in REM mode
- Open MassLynx, and then select the Instrument tab.
- Click MS Method.
- Select REM in the Analyser Mode drop-down menu.
- Ensure that the Lock Mass to be used is appropriate for the REM mass range selected.
NOTE: The precursor mass of leucine enkephalin (m/z 556.27658 / 554.26202) and can be used for the 200-600 m/z and 300-1000 m/z mass ranges whereas the following fragment ion of leucine enkephalin can be used for the 100-300 m/z mass range:
Positive ion (M+H)+ = C11H13N2O3 = 221.09207 (Neutral = C11H12N2O3)
Negative ion (M-H)- = C11H14N3O3 = 236.10406 (Neutral = C11H15N3O3)
- Click Save.
- Select the appropriate ms method files for sample analysis.