Skip to main content
Waters

Calibrating in REM Mode

Article number: 269358

To Calibrate the MRT in REM Mode

From the Tune page:

  1. Select Tune.
  2. Select the required mass range.

clipboard_e7aadda0f66932c7f718fea79dc97b6b4.png

  1. Infuse the calibrantion solution of choice into the MRT, and then ensure that the reference ions are observed.

NOTE:    When choosing a calibration compound, a large number of reference ions across the calibration profile m/z range will achieve greatest mass accuracy. The upper and lower m/z should be also be as close to the m/z range limits as possible. i.e. in the screenshot shown, greatest mass accuracy will be achieved for ions falling within the 248 to 588 m/z range.

  1. Click Acquire to calibrate, in this case sodium formate, over the selected mass range.
  2. Click abort once the acquisition has finished.

clipboard_e960df9e06d33ffbff338c38dceac9179.png

  1. In the Acquisition Settings window, enter a filename and Acquisition time for the mass calibration, and then select Save to acquire the data to be used to perform mass calibration.

IMPORTANT: Ensure that you select the "Retain Data for Mass Calibration" tick box.

clipboard_e6f42a90aa8666610495baa4fb8413634.png

  1. Select manual calibration.

NOTE: Before acquiring the mass calibration, click Abort to ensure that the instrument is not in Tune or acquiring data.

clipboard_e23c600efe5a31ef7744904237110903f.png

clipboard_e7bc094ab08d96567d729581e6ac639fd.png

  1. Select the required REM mass calibration range.
  2. Click Create.

clipboard_e4b41d60f1bb32a29f763835bc078ec29.png

clipboard_e4e26c21eefdf2eafac6c9d6d29d5e206.png

  1. Check the data residuals (Mean prediction error <0.2 ppm), and then click Accept if acceptable.

clipboard_ea2f7d79454619c6ca81309230cc746c2.png

NOTE: A successful calibration over the required mass range and polarity will be indicated.

clipboard_e4732047f94febcbcccd8aa4c41e180ad.png

 

Not able to find a solution? Click here to request help.

  • Was this article helpful?