Calibrating in REM Mode
Article number: 269358
To Calibrate the MRT in REM Mode
From the Tune page:
- Select Tune.
- Select the required mass range.
- Infuse the calibrantion solution of choice into the MRT, and then ensure that the reference ions are observed.
NOTE: When choosing a calibration compound, a large number of reference ions across the calibration profile m/z range will achieve greatest mass accuracy. The upper and lower m/z should be also be as close to the m/z range limits as possible. i.e. in the screenshot shown, greatest mass accuracy will be achieved for ions falling within the 248 to 588 m/z range.
- Click Acquire to calibrate, in this case sodium formate, over the selected mass range.
- Click abort once the acquisition has finished.
- In the Acquisition Settings window, enter a filename and Acquisition time for the mass calibration, and then select Save to acquire the data to be used to perform mass calibration.
IMPORTANT: Ensure that you select the "Retain Data for Mass Calibration" tick box.
- Select manual calibration.
NOTE: Before acquiring the mass calibration, click Abort to ensure that the instrument is not in Tune or acquiring data.
- Select the required REM mass calibration range.
- Click Create.
- Check the data residuals (Mean prediction error <0.2 ppm), and then click Accept if acceptable.
NOTE: A successful calibration over the required mass range and polarity will be indicated.