2545 BGM/QGM fails static leak test - WKB82500

Article number: 82500

SYMPTOMS

  • Pressure fluctuation
  • Static leak test failure

ENVIRONMENT

  • 2535
  • 2545
  • 2555

CAUSE

Check valve failure

FIX or WORKAROUND

Replace the check valve. 

ADDITIONAL INFORMATION

700001493 - Check Valve Cartridge Kit - 2535/2545

700004867 - Check Valve Cartridge Kit - 2555